rss
J Clin Pathol doi:10.1136/jcp.2007.053850

A quality assurance exercise to evaluate the accuracy and reproducibility of CISH for HER2 analysis in breast cancer

  1. Silvana Di Palma (silvana.dipalma{at}royalsurrey.nhs.uk)
  1. The Royal Surrey County Hospital, United Kingdom
    1. Nadine Collins (ncollins{at}royalsurrey.nhs.uk)
    1. Royal Surrey County Hospital, United Kingdom
      1. Michael Bilous (michaelb{at}icpmr.wsahs.nsw.gov.au)
      1. Westmead Hospital, Australia
        1. Anna Sapino (anna.sapino{at}unito.it)
        1. University of Torino, Italy
          1. Marcella Mottolese (sdipalma{at}royalsurrey.nhs.uk)
          1. Regina Elena Cancer institute, Italy
            1. Nikiforos Kapranos (nkapranos{at}mitera.gr)
            1. MITERA Maternity and Surgical Center, Greece
              1. Fernando C Schmitt (fernando.schmitt{at}ipatimup.pt)
              1. IPATIMUP and Medical Faculty of Porto, Portugal
                1. Jorma Isola (jorma.isola{at}uta.fi)
                1. University of Tampere, Finland
                  • Published Online First 6 March 2008

                  Abstract

                  Abstract

                  Aims - Chromogenic in situ hybridisation (CISH) is an alternative to IHC or FISH for the assessment of HER2 oncogene status in breast cancer. Although CISH is being used increasingly in routine diagnostics, there are no established inter-laboratory quality assurance programmes for this test. We wished to assess our performance in carrying out and interpreting HER2 CISH when compared to other centres that were also experienced in the assay.

                  Methods - We assessed the reproducibility of HER2 CISH analysis when performed by seven different centres which use the test routinely in diagnostic service.

                  Results - The results from 28 cases showed overall concordance of 98.5% (192 / 195 tests) (kappa coefficient 0.91). One of the discrepancies was due to the invasive carcinoma having been cut out in the sections received by two of the centres, and the other two were in the non-amplified / equivocal / low amplified category.

                  Conclusion - This is the first report of a quality assurance study assessing laboratories that use HER2 CISH routinely in clinical diagnostics. The results show that CISH is a robust technique providing a suitable assay for the frontline testing of HER2 status in breast cancer.

                  Latest from JCP Education

                  Latest from JCP Education

                  Register for free content


                  Free sample
                  This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of JCP.
                  View free sample issue >>

                  Free archive
                  The full back archive is now available for JCP. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
                  Register to access the free archive >>

                  Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

                1. Latest Pathology jobs

                  Latest Pathology jobs