J Clin Pathol 2005;58:951-954 doi:10.1136/jcp.2004.025429
  • Original article

Can urine dipstick testing for urinary tract infection at point of care reduce laboratory workload?

  1. H D Patel,
  2. S A Livsey,
  3. R A Swann,
  4. S S Bukhari
  1. Department of Clinical Microbiology, University Hospitals of Leicester NHS Trust, Leicester LE1 5WW, UK
  1. Correspondence to:
 Dr S S Bukhari
 Sandringham Building, Floor 5, Leicester Royal Infirmary, Infirmary Square, Leicester LE1 5WW, UK; sayed.bukhariuhl-tr.nhs.uk
  • Accepted 14 January 2005

Abstract

Aim: The University Hospitals of Leicester NHS Trust microbiology laboratory receives 150 000 urine samples each year, approximately 80% of which prove to be culture negative. The aim of this study was to reduce the proportion of culture negative urines arriving in the laboratory, by producing local evidence based guidelines for the use of urine dipstick testing at point of care within the trust’s three acute hospitals.

Methods: One thousand and seventy six unborated urine samples were dipstick tested at the point of care using an automatic strip reader. Quantitative results for the four infection associated markers—leucocyte esterase, nitrite, blood, and protein—were compared with the results of conventional laboratory microscopy and culture.

Results: The performance of different marker combinations was calculated against the routine laboratory methods. One hundred and seventy five (16.3%) samples were negative for all four markers. Of these dipstick negative samples, only three (1.7% of all true positives) were positive by culture. The absence of all four infection associated markers was found to have a greater than 98% negative predictive value and a sensitivity and specificity of 98.3% and 19.2%, respectively.

Conclusions: A urinary dipstick testing algorithm for infection associated markers was derived for use in hospital patients to screen out negative urines. Two years after distributing the algorithm and promoting access to reagent strips and strip readers, a reduction in the urine workload has been seen against an otherwise increasing laboratory specimen load.

Footnotes

    Latest from JCP Education

    Latest from JCP Education

    Register for free content


    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of JCP.
    View free sample issue >>

    Free archive
    The full back archive is now available for JCP. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

  • Latest Pathology jobs

    Latest Pathology jobs