rss
J Clin Pathol 56:114-119 doi:10.1136/jcp.56.2.114
  • Original article

European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors

Table 4

Monitor and reference INR for 535 target samples at 10 centres

POCT system Species specific IRP Mean reference INR Lyophilised plasma calibration Mean monitor INR Mean % difference CI for mean % difference p Value (paired t test)
Geometric means of reference and monitor INR values, percentage difference in geometric mean INR, corresponding 95% confidence intervals (CI), and p value are given.
INR, international normalised ratio; IRP, international reference preparation; POCT, point of care testing; RBT/90, rabbit plain international reference preparation; rTF/95, human recombinant international reference preparation.
CoaguChek Mini RBT/90 2.50 Artificially depleted 2.77 10.8 8.9 to 12.7 <0.0001
Individual coumarin 2.56 2.2 0.5 to 3.9 0.0099
Pooled coumarin 2.63 5.0 3.3 to 6.8 <0.0001
TAS PT-NC rTF/95 2.46 Artificially depleted 2.16 −12.0 −13.2 to −10.8 <0.0001
Individual coumarin 2.13 −13.4 −14.5 to −12.2 <0.0001
Pooled coumarin 2.14 −12.9 −14.1 to −11.7 <0.0001