J Clin Pathol 2002;55:467-471
  • Technical report

Comparison of Ki-67 equivalent antibodies

  1. C F Lindboe1,
  2. S H Torp2
  1. 1Department of Pathology, Vest-Agder Central Hospital, N-4604 Kristiansand, Norway
  2. 2Department of Laboratory Medicine and Pathology, Norwegian University of Science and Technology, Trondheim University Hospital, N-7006 Trondheim, Norway
  1. Correspondence to:
 Dr C F Lindboe, Department of Pathology, Vest-Agder Central Hospital, N-4604 Kristiansand, Norway;
 c.lindboe{at}c2i.net
  • Accepted 19 September 2001

Abstract

Aims: To compare commercially available Ki-67 equivalent antibodies with regard to qualitative and quantitative immunohistochemical staining characteristics.

Methods: The following antibodies were used: monoclonal MIB-1 (Immunotech), monoclonal MM1 (Novocastra), polyclonal NCL-Ki-67p (Novocastra), and polyclonal Rah Ki-67 (Dako). All immunostainings were evaluated in squamous epithelium from formalin fixed and paraffin wax embedded pharyngeal tonsils. Labelling indices (LIs) were recorded twice to test their reproducibility.

Results: By application of all four antibodies the nuclear staining could be either diffuse, granular, or a combination of both (classified as granular in this study). The diffuse pattern generally showed a strong or moderate staining intensity, whereas the granular pattern displayed a continuum from strong to very weak, making it difficult to discriminate between positive and negative nuclei. The diffuse staining pattern was seen in approximately 59% of the nuclei with the MIB-1 antibody and in 35–45% when the other antibodies were used. The following mean LIs were recorded: MIB-1, 31%; NCL-Ki-67p, 21%; Rah Ki-67, 17%; and MM1, 14%. The reproducibility was excellent for all four antibodies, with the mean of differences between the two runs of counts ranging from 1.1% to 1.5%.

Conclusions: The four tested Ki-67 equivalent antibodies revealed differences in qualitative and quantitative staining characteristics, which resulted in considerable variations in registered LIs. The MIB-1 antibody appears to have a higher sensitivity for detecting the Ki-67 antigen than the other three tested antibodies. These differences are important to consider when proliferative activity is determined by the Ki-67 LI.

Footnotes

    Latest from JCP Education

    Latest from JCP Education

    Register for free content


    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of JCP.
    View free sample issue >>

    Free archive
    The full back archive is now available for JCP. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

  • Latest Pathology jobs

    Latest Pathology jobs