J Clin Pathol 2001;54:25-30 doi:10.1136/jcp.54.1.25

An evaluation of three commercial kits for use as screening methods for the detection of leptospiral antibodies in the UK

  1. W J Zochowski1,
  2. M F Palmer1,
  3. T J Coleman1
  1. 1PHLS Leptospira Reference Unit/WHO/FAO Collaborating Centre for Research on Leptospirosis, County Hospital, Hereford HR1 2ER, UK
  1. Mrs Zochowskiwendy.zochowski{at}talk21.com
  • Accepted 20 July 2000

Abstract

Aims—To compare three commercial screening tests—the PanBio leptospiral IgM enzyme linked immunosorbent assay (ELISA), the Biolisa leptospiral IgM ELISA, and the indirect haemagglutination assay (IHA)—with the microscopic agglutination test (MAT) and two “in house” ELISAs—urease and horseradish peroxidase (HRP)—for the detection of leptospiral antibodies in a local UK and Eire population.

Method—Two hundred sera submitted for a differential diagnosis of leptospirosis were tested by all methods. A further 142 sera from patients with antibodies to toxoplasma, Epstein-Barr virus (EBV), hepatitis A virus, rheumatoid factor, Borrelia burgdorferi, Mycoplasma pneumoniae, syphilis, cytomegalovirus, and Q fever were tested for crossreactivity.

Results—Compared with the MAT, sensitivity and specificity were found to be: PanBio, 90%/94%; Biolisa with sorbent, 100%/85%; and IHA, 54%/95%. Seven of 200 trial sera gave false negative results with PanBio; 14 of 200 trial sera gave false positive results with Biolisa with sorbent, as did a further 25 of the 142 sera tested for potential crossreactivity. Two of 142 sera gave crossreactions with PanBio and IHA (one each).

Conclusions—The degree of false positivity seen with the Biolisa suggests that the recommended positive value of ≥ 26 Eu/ml should be reassessed using pools of sera from local populations. When the cut off value was reassessed, using a value of ≥ 40 Eu/ml, a sensitivity and specificity of 96% and 94%, respectively, was achieved. Even the modified Biolisa appears to be over sensitive and to show a high degree of non-specificity. The IHA, although specific (95%), lacked sensitivity in this study. The PanBio appeared to be the most suitable as a screening test for leptospiral IgM in the UK, although it would be advisable for all positive test results to be confirmed by a different enzyme immunoassay and the MAT.

Footnotes

    This Article

    Services

    1. Request permissions

    Responses

    1. Submit a response
    2. No responses published

    Social bookmarking

    Latest from JCP Education

    Latest from JCP Education

    Register for free content


    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of JCP.
    View free sample issue >>

    Free archive
    The full back archive is now available for JCP. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

  • Latest Pathology jobs

    Latest Pathology jobs