rss
J Clin Pathol 53:929-932 doi:10.1136/jcp.53.12.929

Audit and internal quality control in immunohistochemistry

Table 4

Scores in percentages of all test slides reviewed in our study, including positive tissue controls where the test case slide was negative

Scores for each year
0 1 2 3
Staining criteria 1996 1997 1998 1996 1997 1998 1996 1997 1998 1996 1997 1998
Total slides reviewed by year (manual methodology only): 1996, 119; 1997, 90; 1998, 72.
NA, not applicable.
Staining intensity 0.0 0.0 0.0 0.8 1.1 0.0 0.0 8.9 8.3 99.2 90.0 91.7
Uniformity of staining 5.0 4.4 0.0 95.0 95.6 100.0 NA NA NA NA NA NA
Specificity of staining 0.8 6.7 2.8 99.2 93.3 97.2 NA NA NA NA NA NA
Absence of background staining 0.3 1.1 0.0 36.1 17.8 38.9 63.6 81.1 61.1 NA NA NA
Counterstaining 0.0 1.1 0.0 100.0 98.9 100.0 NA NA NA NA NA NA