rss
J Clin Pathol 1988;41:397-402 doi:10.1136/jcp.41.4.397
  • Research Article

Laboratory assessment of three reflectance meters designed for self monitoring of blood glucose concentrations.

  1. J P Begley,
  2. A R Forrest
  1. Department of Clinical Chemistry, Royal Hallamshire Hospital, Sheffield.

      Abstract

      Three monitors designed for self monitoring of blood glucose concentrations by diabetic patients were evaluated in a laboratory. All three assessments correlated positively with a laboratory reference method based on glucose oxidase. Coefficients of variation at all levels tested were less than 10% for each monitor. Stability of the colour development of each stick was assessed and the effects of changes in blood spot volume and incubation time were examined. Twelve ward nurses were each asked to measure two blood samples and the results obtained showed wide variability. Overall, our findings suggest that each of the three monitors tested is suitable for use in monitoring capillary blood glucose concentrations by those who are properly trained.

      Latest from JCP Education

      Latest from JCP Education

      Register for free content


      Free sample
      This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of JCP.
      View free sample issue >>

      Free archive
      The full back archive is now available for JCP. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
      Register to access the free archive >>

      Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

    • Latest Pathology jobs

      Latest Pathology jobs