rss
J Clin Pathol 1980;33:243-248 doi:10.1136/jcp.33.3.243
  • Research Article

Interlaboratory comparison of serum vitamin B12 assay.

Abstract

The results have been compared of microbiological and radioisotope dilution (RID) assay of serum vitamin B12 by participants in national interlaboratory trials in Britain. There was wide variation between the individual participants, especially marked in the L. leichmannii microbiological assay and in the RID methods, whereas excellent correlation, reproducibility, and recovery were obtained in reference laboratories by microbiological assay with both E. gracilis and L. leichmannii. In general, RID gave higher results than microbiological assay. The need for suitable reference sera is emphasised.

Latest from JCP Education

Latest from JCP Education

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of JCP.
View free sample issue >>

Free archive
The full back archive is now available for JCP. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.